EZ Match User Manual
Initiating Calibration
Calibration Progress Output to Terminal
When initiated by serial command or EZ Button, the calibration operation will output
progress and process information to a terminal if the system is configured to do so.
Terminal
Depending upon how the calibration options are configured, the calibration process can
have up to 5 steps (Search Pass, Medium Pass, Fine-Tune Pass). Calibration data is
shown in a simple table format under 8 category headings, each of which are defined
below.
Calibration Progress Output Examples (Terminal)
Search Pass
The purpose of Search Pass is to decode a symbol in as few configurations as possible
during calibration. This process is designed to identify symbology type, and to identify any
other image processing parameters that are already calibrated.
Heading Definition
Prog Progress indicator with a range of 0 - 100. Indicates percent complete.
Rating Rating indicator; higher ratings are more favorable.
Decode Number of successful decodes for the pass.
Quality Quality indicator; higher ratings are more favorable.
Shutter Shutter setting under operation.
Gain Gain setting under operation.
Locate Time required to locate symbol, expressed in milliseconds.
Running Search Pass
Prog Rating Decode Quality Focus Shutter Gain Locate
12 0 0 0 313 250 0 0
14 0 0 0 313 250 9 0
16 0 0 0 313 250 18 0
18 2174 1 72 313 250 27 24