QX Hawk Industrial Imager User Manual 7-29
Utilities
Static Validation
Sending a <VAL> command from ESP’s Terminal will cause the imager to perform a static
validation operation on a Data Matrix symbol and will output validation results to the Terminal.
Output Screen
Data Matrix validation information is output in the format below in response to the <VAL>
command.
Note:
If the imager fails to decode the Data Matrix symbol, a No Read message is displayed.
Validation Parameters
Symbol Contrast
Usage: Lets the user know if contrast settings are less than acceptable.
Definition: All the pixels that fall within the area of the test symbol, including its
required zone, will be sorted by their reflectance values to select the
darkest 10% and the lightest 10% of the pixels. The arithmetic mean of
the darkest and the lightest pixels is calculated and the difference of the
two means is the Symbol Contrast.
(ANSI) Symbol Contrast grading is defined this way:
A (4.0) if SC ? 70%
B (3.0) if SC ? 55%
C (2.0) if SC ? 40%
D (1.0) if SC ? 20%
F (0.0) if SC ? 20%